*Specifications:
- Test Frequency: 1 MHz - 3 GHz with 100 kHz resolution. With list-sweep, up to 32 points per sweep is available.
- Impedance Parameters: |Z|, ¥èz (rad), ¥èz (deg), |Y|, ¥èy (rad), ¥èy (deg), X, G, B, Ls, Lp, Cs, Cp, Rs, Rp, Q, D
- Display Resolutions: 5 digits
- Test Signal Level: V (open condition): 4.47 mVrms - 502 mVrms (447 mVrms when >1 GHz) I (short condition): 0.0894 mArms - 10 mArms (8.94 mArms when >1 GHz)
- Basic Accuracy: ¡¾1.0%
- Measurement Range: 200 m¥Ø to 3 k¥Ø (@1 MHz, accuracy ¡Â10%)
- Measurement Time: 9 msec per point (max. speed)
- Measurement Terminal: 3.5 mm (female)
- Calibration and Compensation: Open/short/load/low-loss capacitor calibration, fixture electrical length compensation, open/short compensation
- Rdc Measurement Function: For contact check (on/off selectable)
- Data Storage Devices: About 18 GB internal hard disk and 1.44 MB floppy disk
- Interface: GPIB, LAN (10base-T/100base-TX automatically switched), and Opto-isolated handler interface
- Includes: Test Head with 1 m Cable, 3 each N(m) to SMA(f) adapters